FloTrac: pulse contour analysis

This chapter is relevant to Section G6(iii) of the 2023 CICM Primary Syllabus, which expects the exam candidate to "describe the methods of measurement of cardiac output, including limitations, potential sources of error, the need for calibration and the values obtained".  Question 4 from the second Fellowship Exam paper of 2024 was the only time FloTrac had ever appeared in the exams, and even then they asked the candidates to choose one from PiCCO, LiDCO and FloTrac which means one may have skirted around the need to know about it. To be precise, FloTrac™ is the sensor and Vigileo™ is the monitor device marketed by Edwards, and both are required to make measurements.

Principles of cardiac output monitoring by FloTrac

  • The FloTrac monitor is an uncalibrated pulse contour analysis monitor
  • This works on the premise that, if the compliance of the arterial circulation is known, then the stroke volume can be calculated from the pulse pressure, because that change in pressure must have resulted from a corresponding change in volume.
  • Other methods like PiCCO and LiDCO use thermodilution or indicator dilution to assess the compliance of the circulation, but these tend to drift in accuracy over time as the compliance value changes
  • Instead of this, the FloTrac monitor estimates the compliance on the basis of the waveform contour and some loookup tables from empirical measurements, i.e. it does not require calibration by the user -calibrating data are derived from patient demographics
  • The algorithm used to perform those estimates is proprietary

Advantages of FloTrac

  • No need to calibrate - therefore, instantly useable from insertion
  • Only requires an arterial access point, no need for a central line
  • The technique shows good agreement with PA catheter thermodilution measurement.
  • No additional data beyond patient demographics is needed
  • No indicator injectate - easier operation
  • No need to sample the blood, and therefore no iatrogenic blood loss
  • Not confused by shunts or valvular regurgitation
  • No calibration, means no drift from calibration

Limitations

  • Made inaccurate by atrial fibrillation, ECMO or IABP
  • The presence of wildly deranged systemic vascular resistance values will reduce the validity of the cardiac output measurements
  • It literally only does the cardiac output stroke volume variation and SVR, all other sophisticated variables (PA pressure, wedge pressure, GEDV, etc) are not available
  • The patient population in the data set used to generate the data may not be representative of the patient being monitored, eg. it does not cover severe sepsiss, liver failure, etc