Basic features of the ECMO circuit

ECMO is half of "Mechanical supports: ECMO/IABP", an L1 topic of Section 2.1.4 in the second edition of the CICM Syllabus for the Second Part Examination. This chapter is a catch-all category for fundamental ECMO material which is simultaneously too basic to be asked in the CICM Second Part Exam, and too complicated to be asked in the CICM First Part Exam, but completly fair to be asked by wide-eyed juniors to their knowledgeable senior supervisors. Questions such as "how is this thing different from the bypass in theatres" or "what is this thing on the side of the circuit tubing". The author has been that senior supervisor, and the answers he has previously given had been so glib and underwhelming, that the construction of this section seemed like a self-education imperative. 

There are no journal publications that specifically address the material presented here in a way that would make them a clear recommendation to the time-poor reader, which means that the best resource for all things ECMO is the series of ECMO management guidelines from Man's Best ICU. This is a good balance of succinct and comprehensive, and can be regarded as a peer-reviewed resource, as it was written with the collaboration of multiple senior CICM fellows, among whom were several college examiners. On the other hand, for the reader with abundant time, Chris Nickson's ECMO section from Intensive is an excellent one-stop solution to the prolix with which this website is so markedly affected. Moreover it acts as a central hub of references to other professional resources. Specifically for machinery, this Cardiohelp service manual from Maquet is also very helpful.  

How is this not cardiopulmonary bypass

ECMO, a motorised outboard circulatory system, clearly resembles the cardiopulmonary bypass

machine used to stop the heart and pause the natural process of death while the patient undergoes cardiac surgery. This begs the question, if these systems do the same sort of thing, then why do we have both? Is that not such a wasteful excess of devices? The answer, of course, is that they are far

from interchangeable.  Sinard & Bartlett (1990), writing at a time when this was a novel concept, summarise the main technological differences:

  • Stagnant pools are avoided. This is the main difference: there are no venous reservoirs. The bypass circuit drains blood from the patient and puts it into a, for lack of a better word, bucket. This bucket acts as the sink for all the blood drained from the cardiac chambers, the venous circulation, and various suckers; the blood here is defoamed, filtered and deaerated before being returned to the patient. This might sound like an unnecessary additional step, but in fact it serves the purpose of acting as a buffer of blood for the pump. Were the venous drainage suddenly to run dry (eg. where the surgeon did something dramatic), the venous reservoir will continue to supply blood to the patient, giving the perfusionist and anaesthetist some reaction time to respond to the situation. As the result, there is usually at least about 1000ml in there, which gives you about 15-20 stroke volumes, translating into about 10 seconds or so for some frantic activity.
    The venous reservoir receives blood, which is commonly drained by gravity alone, and so raising or lowering the level of the reservoir effectively controls the blood volume of the patient; or it can be under gentle suction to allow for a rapid rate of venous return to the pump. This is a luxury of temporarily having blood that is not inclined to clot, but it is not a good long term option, as the resulting pools of blood tend to create a stagnant environment that encourages platelet degranulation and aggregation.
  • Blood never contacts air directly. The venous reservoir contains air-blood interfaces which are intensely immunogenic, as they tend to activate neutrophils and trigger platelet activation (Lesouhaitier et al , 2024). In fact, when Mulholland et al (2000) tested different circuit components, they found that the combination of negative pressure and air/blood interface (of the sort that exists in the hardshell reservoir) created about fifteen times as much free haemoglobin as merely rubbing against PVC circuit components.
  • The circuit contains a rather fixed volume of blood.  The patient is the reservoir, and the ECMO circuit does not vary in volume. 
  • Cannulation is peripheral. The cannulae, perfusion (return) and siphon (access), are usually in the large peripheral veins.
  • There is an option to only replace the lungs, whereas straight CPB almost by definition does both the heart and the lungs.
  • There are no filters or de-aerator traps, apart from the oxygenator, which is mostly because the ECMO circuit will never need to receive blood  aspirated from the surgical field.
  • The pump is a centrifugal rotor, whereas  many bypass machines still use peristaltic roller pumps (replacing the single-use silicone tubing is cheaper than replacing a big expensive maglev rotor, and the rollers are less sensitive to afterload)
  • There is no cardioplegia equipment, as there is no intentional plan to stop the heart on ECMO.
  • The patient is not undergoing surgery. This is a critical issue. Certainly the surgical injury from diathermy retraction and incisions contributes considerably to the inflammatory sequelae of CPB, but the most important component is probably the reperfusion of the heart and lungs which probably contributes the majority of the inflammatory insult following bypass.

The upshot of these differences is:

  • Cannulation is easier to perform
  • Cannulas are easier to tolerate
  • Circuit configuration is much simpler 
  • The circuit does not have to be anticoagulated as much as for CPB
  • The circuit can run for days or weeks, because it is less immunogenic
  • The amount of circuit surface which the blood is exposed to is slightly smaller, and the opportunity to stagnate and form clots is much smaller, which means the immunological inflammatory and haematological consequences of ECMO are much more gentle than "proper" bypass
  • The total circulating volume is expanded by a smaller amount (the volume of the ECMO circuit is much smaller, perhaps 500ml in total, and as low as 150ml in some configurations, whereas the bypass circuit could be up to 2L, depending on how much blood is in the venous reservoir). As an example, the author's local workplace carries Cardiohelp HLS circuits with an average priming volume of 570ml for a total tube length of 4.6m.

The ECMO circuit in simplest summary

Fighting temptation to draw and contrast the ECMO and CPB diagrams, the author will cling to some authenticity and authority by only describing what he himself has direct experience of, having never been a cardiac anaesthetist or surgeon. ECMO circuits are thankfully simple, and can be described as a simple loop of blood filled tubing connecting a patient to an oxygenator. The flow of blood needs to be maintained in some way, and a pump is used to provide that energy, which makes the simplest ECMO circuit boil down to a diagram like this: 

The simplest possible diagram of an ECMO circuit

This is obviously a gross oversimplification, as it leaves out a whole host of important circuit components, but is a good start to understanding how the patient's blood is accessed, circulated through a gas exchange device, and then returned to the patient. "Access" and "return" become embedded in the reader's thinking by this method, and this is important because the nomenclature of ECMO is occasionally a barrier to understanding the techniques. 

A more detailed examination of that circuit can be attempted, using a VV configuration as an example:

ECMO circuit diagram with labels of the circuit components

This diagram has proven very useful in the course of doing approximately one million powerpoint presentations about ECMO, and for those who are interested, a blank image of an unlabelled central circulatory system to help create your own ECMO diagrams is left here as a download, as well as the original Adobe Illustrator vector file used to put it together. Of course this is quite pointless to those of us who are either a) kinesthetic learners or b) not inclined to waste their time in an exam drawing elaborate diagrams of the circulatory system. In which case, the ECMO circuit can be represented by the following unordered list:

  • Access cannula
  • Circuit tubing
  • SvO2 sensor
  • Pre-pump  pressure sensor
  • Pump
  • Post pump pressure sensor
  • Oxygenator
    • Water heater
    • Gas blender
    • Dialysis access and return ports
  • Post oxygenator pressure sensor
  • Flow sensor and bubble alarm
  • Return cannula

This list of components became the defacto table of contents for this section, which mostly deals with the technology of the different parts of the circuit. However, acknowledging that not everybody is interested in the adsorptive properties of polymerised vinyl, the following short summary of circuit elements is offered, giving each element some sort of personality by either linking to a page about their clinical relevance or mentioning some memorable "superpower" of that component module.

A brief tour of ECMO circuit components

  • Access cannula is usually a "multi-stage" (i.e. many holes) large bore device in a large venous structure or even the right atrium. These can be up to 32 Fr on the French gauge system, i.e. an outer diameter of 10.67 mm - for many of us, the largest object one has ever introduced into a patient. 
  • Circuit tubing  is usually made of PVC doped with plasticisers and various other distinctive property enhancers which make it suitable for its purpose (it has to be transparent, flexible, robust, kink-resistant, biologically inert, etc etc). The convention is to use tubing 3/8th inch in internal diameter. This tubing has the distinctive superpower of being extremely absorptive and ADsorptive, becoming gradually encrusted and infused with molecules originating from the patient (for example, midazolam). 
  • SvO2 sensor is sufficiently unlike the usual SpO2 sensor that the two cannot be used interchangeably. It is more like the forehead SpO2 sensor, which uses reflection spectrophotometry to assess the oxygenation of haemoglobin. Unlike the finger or forehead, the venous tube has no pulse, and this would confuse devices which are looking for a pulse (conversely, pulsatile flow in a human digit would bewilder the ECMO sensor). This sensor offers a convenient early warning of falling cardiac output, tissue hypoxia, and recirculation.
  • Pressure sensors are depicted as being positioned around the circuit, but this is in fact an artifice to enhance the explanatory power of the diagram. In reality many modern systems integrate these into the combined pump/oxygenator/heat exchanger consumable item, making the exterior sensors obsolete. Where such exterioir sensors exist, they are Wheatstone bridge transducers that connect to a backpressure bag similar to the arterial line set, and can even be used for sampling. Where they are integrated, they are also Wheatstone transducers, except the deformable silicone gel/polysulfone membrane is exposed directly to the bloodstream in the oxygenator module.
  • Pre-pump  pressure sensor measures the Pven, the pressure of the access line, and is usually faily negative (around -30 mmHg is normal, and anything close to -100 mmHg is increasingly concerning)
  • Pump is a centrifugal pump which is powered by a magnetically coupled motor. In function, it resembles the magnetic stirrer one may recall from tedious chemistry practicals, which use a set of stationary electromagnets with rapidly alternating current is used to push a ferromagnetic stir bar ("flea", "spinner", etc) in order to mix the contents of the beaker. The pump drive head is suspended magnetically within the chamber and makes no contact with its walls, limiting the sources of red cell destruction to mostly just shear stress. Its function is measured ias rotational velocity, as rpm; and a normal pump speed is around 2500-3000 rpm. Anything in excess of 4000 is likely to produce haemolysis. The motor behind the pump is a 12-28V DC motor capable of outputting about 50 Watts per hour, approximately equivalent to a small handheld car vacuum or a cordless rotary tool.
  • Post pump pressure sensor measures the Pint, which is the pressure generated by the pump and exerted on the hollow fibres of the oxygenator. This value is usually the highest value in the house, and is usually around 150-200 mmHg, depending on the pump speed (the high pressure alarm defaults to a value of 400 mmHg, which gives you an idea of the range)
  • Oxygenator is the integrated hollow fibre assembly that allows blood, gas and water to parallel-play together and politely exchange their desirable properties while keeping their hands to themselves. These vary in design and properties so much that one cannot make generalisable statements about them, but the general range seems to be between 0.8 m2 and 2.5 m2.
    • Water heater is usually a simple kettle with a pump and thermostat, functionally indistinguishable from enthusiast-grade aquarium equipment. The local HU-35 is capable of delivering a maximum of 500W of heat, and is said to pump with a 10 W pump at a maximum flow rate of 5L/min, which makes it suitable for a 350-400L tropical tank full of sensitive Cichlids. 
    • Gas blender is a crude mechanical device composed of machined steel plates and bolted together in a manner which resembles 1970s Soviet cars, a totally analogue steampunk anachronism among the high-tech equipment of the ICU. The local model is supplied by Sechrist, who also manufacture hyperbaric oxygen chambers. The superpower of this device seems to be the sneaky delivery of extra oxygen which occurs even when the dial is turned to "21%", which means one should physically disconnect it from the wall during VV ECMO weaning tests.
    • Dialysis access and return ports are vascath-sized openings which allow a CRRT circuit to be connected inline with the ECMO circuit. Oxygenated blood is removed from the post-oxygenator circuit, pumped through the CRRT machine, and returned to pre-oxygenator post-pump line. This curious feature gives rise to the otherwise never-seen "Return pressure extremely positive" alarm in the CRRT circuit, as the poor dialyser spins its little peristaltic rollers to pump against the pressure generated by the roaring ECMO impeller.
  • Post oxygenator pressure sensor is integrated to assess the pressure in the return limb of the circuit (Part) and is the first to notice that the return cannula is kinked or disconnected. It is also used to calculate the ΔP, pressure across the oxygenator, which is an indicator of the oxygenator health. 
  • Flow sensor and bubble alarm are often integrated into the same module, and use an ultrasonic detector to Doppler the blood as it passes through them. A side effect of using this technology is the ability to easily detect small air bubbles. These devices are usually clamped to the return line of the circuit to look for bubbles, and to the femoral perfusion line for VA ECMO to assure that the leg is getting some bloodflow.
  • Return cannula is usually smaller than the access cannula (commonly 19Fr), mostly because the smallest size is selected to deliver the appropriate flow. Wickramarachchi et al (2025) experimented with a number of them and found that the practical limit is probvably something like 13Fr, as this delivers only 3L/min of return blood with a Part of around 400 mmHg (at the limit of what the sensors will tolerate).

References

Molyneux, Victoria, and Andrew A. Klein. "Equipment and monitoring for cardiopulmonary bypass." Cardiopulmonary bypass 2 (2022).

Sinard, J. M., and Robert H. Bartlett. "Review articles: Extracorporeal membrane oxygenation (ECMO): prolonged bedside cardiopulmonary bypass." Perfusion 5.4 (1990): 239-249.

KIRBY, CHARLES K., et al. "Simple automatic method of blood volume control during cardiac bypass for open heart surgery." AMA Archives of Surgery 78.2 (1959): 193-196.

Scott, W. F. "Fluid volume control in heart-lung bypass operations in open-heart surgery." South African Medical Journal 34.13 (1960): 249-252.

Lesouhaitier, Mathieu, Félicie Belicard, and Jean-Marc Tadié. "Cardiopulmonary bypass and VA-ECMO induced immune dysfunction: common features and differences, a narrative review." Critical Care 28.1 (2024): 300.

Callaghan, Sam, et al. "Adsorption of blood components to extracorporeal membrane oxygenation (ECMO) surfaces in humans: a systematic review." Journal of Clinical Medicine 9.10 (2020): 3272.

Wickramarachchi, Avishka, et al. "Flow capabilities of arterial and drainage cannulae during venoarterial extracorporeal membrane oxygenation: A simulation model." Perfusion 40.3 (2025): 668-677.

Mulholland, J. W., W. Massey, and J. C. Shelton. "Investigation and quantification of the blood trauma caused by the combined dynamic forces experienced during cardiopulmonary bypass." Perfusion 15.6 (2000): 485-494

Boettcher, Wolfgang, Frank Merkle, and Heinz-Hermann Weitkemper. "History of extracorporeal circulation: the invention and modification of blood pumps." The Journal of ExtraCorporeal Technology 35.3 (2003): 184-191.