ECMO is half of "Mechanical supports: ECMO/IABP", an L1 topic of Section 2.1.4 in the second edition of the CICM Syllabus for the Second Part Examination. And ECMO sensors are 100% of the machine alarms, responding to which is a major part of being a responsible ECMO keeper.
In summary,
The excellent article by Ortoleva et al (2021) was used in the interpretation of haemodynamic variables, but for the rest, no single reference can be suggested. All the information was widely scattered and what few nuggets of valuable information could be found had to be carefully extracted out of the deepest filing cabinets in the basements of digital libraries.
The image from the local Cardiohelp device that just happened to be operating on the morning of writing this is hopefully cropped and resized small enough that the readers will not notice the shameful layer of white fluff collecting on the surface (oh, the state of it).

The main reason to include the image is to point out, like a great pedant, that "SvO2" is not what is being measured here. The reader is directed to the chapter on venous oxygen content for a reminder of what that actually is (mixed blood in the pulmonary artery, is the short version). As you can see from the image, this sensor is actually plugged directly into the Cardiohelp module, and samples blood which is just about to enter the oxygenator, making it an SsomethingO2 monitor, where the something is whatever the position of the access cannula tip is. It could be SVC, RA, IVC, or some random fractional combination of all of the above. Yes, it is mixed blood, and it is venous, but there is a reason for why we have treaties to minimise ambiguity in nomenclature. ELSO themselves would prefer for this to be referred to as SpreO2 to indicate that the sampled blood is pre-oxygenator. But "SvO2" is how it is labelled in many of the professional publications, and worst of all on the ECMO device console itself. Therefore, this resource will also continue to use the incorrect terminology, albeit with much cringing and muttering.
Anyway: the captioned diagram below the photograph is from the original Maquet brochure for their Cardiohelp device, and does not reveal much more about the design or function of the sensor, other than to explain that it is "optical". This does the intricacy of this instrument a disservice, as it can measure the oxygenation of haemoglobin, as well as the haematocrit and the temperature of the blood. Nakano et al (2019) describe the challenges of designing such a device, but the device described in their paper is not the same probe as the one on your local ECMO machine, and chances are that you will not find any details about the sensor array of your specific machine. They appear to measure the oxygen content of mixed venous blood like the pulse oximeter, but do not require the light to be transmitted through the blood, suggesting that reflection spectrophotometry is the mechanism underlying their action (like the forehead SpO2 sensor).
This sensor offers a convenient early warning of falling cardiac output, tissue hypoxia, and recirculation.
An SvO2 value in the 65-75% ballpark is accepted as normal, but on close inspection, this is a figure derived mostly by extrapolating with our experience with mixed venous samples. Consider a V-V situation where the venous sample is not diluted by any recirculating blood. This sample is taken from blood that is aspirated predominantly from the proximal holes of the multistage cannula in the IVC, so it really mostly represents lower body blood, but let's also assume that the aspirated blood represents a true democratic representation of the body's cells. In this rare case, it would still be missing the blood from the coronary sinus, and would only be as accurate as the ScVO2. In case the reader may be wondering whether there is much of a difference between "true" mixed values and ECMO drainage specimens, John et al (2024) measured 82 paired samples and found a nonreassuring mean difference of 11.37%. That was on V-A ECMO as well, i.e. recirculation would not have been a very serious concern.
The ECMO device will usually report temperature measurements of various sorts. As this is something far from the mind of the casual ECMO shopper, the companies that manufacture these devices rarely make specific mention of what and where the measurement is taken by. In the Cardiohelp module, this is measured at the same sensor array as the oximetry.
Is it accurate? Yes, and no. The ECMO temperature can generally be assumed to represent the patient's temperature at least as well as the tympanic probe and the IDC thermistor, and moreover the ECMO heater controls the temperature of the blood by a feedback loop, which means that if the patient tries to change their own temperature, the ECMO circuit will not let them. Theoretically, it should be impossible to be febrile while being on ECMO, though realistically, this is often seen in clinical practice. The usual type of ECMO circuit does nothing to actively cool the patient, unless specifically enhanced to do so, and relies on passive heat loss from the circuit to regulate the temperature when the heating element of the water bath is turned off. Considering Mojoli et al (2015) measured a circuit heat dispersion of 58 ± 12 W and the human body produces 167.2 W to change temperature from 37°C to 40°C, one can envision how even a patient on ECMO could develop a fever if properly motivated.
One needs to have pressure sensors, either scattered around a circuit or integrated into it, to report on the health of the system. There are typically three:
There are, therefore, four pressure variables measured in the ECMO circuit, which are usually reported as a part of the other readouts on the operator panel.

These all seem to be Wheatstone transducers. The exterior sensors are sold separately and appear to be standard transducers similar to arterial line sets, and the integrated sensors dange the deformable silicone gel/polysulfone membrane directly in the bloodstream in the oxygenator module. 
This integrated placement is preferred: the further from the pump you get, the more the measurement is diluted by the effect of the resistance of the tubing, its collapse under negative pressure, its distension due to positive pressure, and other factors. Augustin et al (2010) found that pressures measured at the cannula deviated from the pressures measured at the pipe inlet with increasing flow, to the effect that the pressure value nearest to the pump was much higher (-210 mmHg vs - 100 mmHg measured at the venous cannula).
Those values seem somewhat extreme. The usual operating pressure range, to summarise, is:
In the Cardiohelp user manual, the company states that these sensors lose their accuracy and stop reading at pressures lower than -500 and greater than + 900 mmHg, which seems like an extremely generous margin for any device that applies pressures to a living organism. Actual alarms are usually set much lower. Which begs the question: what happens if you exceed these ranges?
The most fragile part of the circuit is the blood. The plastic components of the tubes and oxygenator are sufficiently stern to put up with a remarkable amount of punishment. Exactly how much is a matter of speciulation, as there's not a lot of data on this in the literature. Some authors report on the performance of specific innovations or component pieces (eg. Abada et al, in 2018, reported that their silcone oxygenator membrane was able to tolerate 260 cmHg without rupture, i.e. 2600 mmHg). It is not clear where Lequier et al (2013) got their value from, but they claim that the whole circuit can tolerate up to 600mmHg in either direction, which seems like a lot. One can concieve of a visually spectacular Mythbusters-style experimental confirmation of these values and ranges, but the author is unlikely to be permitted to do this in his home institution (readers are invited to submit images and video of their own detonated ECMO circuits). The bottom line is, the patient's ability to tolerate insane circuit pressures is the main limiting factor.
"Blood trauma" is what happens from high circuit pressures. The term was used by Mulholland et al (2000), whose paper (though mostly about bypass circuits) seems to be quoted by basically everyone who discusses the subject. In short, the combination of negative pressure and air interface was the most damaging, producing about fifteen times more haemolysis than merely contact with circuit surfaces. The next most damaging was negative pressure, followed by positive pressure, and then all the other tested variables. Empirically, the negative pressure necessary to cause haemolysis seems to be around -100 mmHg; below this value it appears to be quite stable, but it increases linearly with increasingly negative pressure beyond this. Positive pressure, on the other hand, seems relatively harmless, apart from sandblasting the aortic atheroma.
One might ask, why is there such a difference between negative and positive pressure? Surely a pressure gradient is a pressure gradient, irrespective of the direction. And that might be reasonable if shear stress was the most important mechanism, but in fact it seems that most of the harm is done by microscopic cavitation and outgassing. The latter is a phenomenon produced by localised areas of negative pressure in turbulent fluid, which permits dissolved gases to briefly appear as buubles, and then to collapse immediately as the conditions revert to normal. Girod et al (2002) is the paper that gets quoted when this topic comes up in the literature; it is an old study looking at microbubble phenomena at the leaflets of mechanical valves (the bubbles were carbon dioxide). They did not detect cavitation per se, which is distinct from outgassing (the bubbles are vapour of the solvent rather than dissolved gas), and which is thought to be more damaging (it is the same phenomenon that erodes the blades of naval propellers).
After having one's faith in the accuracy of circuit SvO2 shaken, the reader will surely find themselves questioning other cherished parameters. How does ECMO affect the numbers we look at from the end of the bed to assess whether the patient is getting better or worse? There are various changes that occur, and none of them reduce the validity of usual monitoring techniques, provided one can adjust their thinking to account for the predictable effect of the ECMO circuit. Consider:
Therefore:
But some things still matter:
Broman, Lars Mikael, et al. "Time to throw out the elephant in the room: Proper use of SvO2 in extracorporeal life support." Asaio Journal 71.1 (2025): e17-e18.
Broman, Lars Mikael, et al. "The ELSO Maastricht Treaty for ECLS Nomenclature: abbreviations for cannulation configuration in extracorporeal life support-a position paper of the Extracorporeal Life Support Organization." Critical Care 23.1 (2019): 36.
Conrad, Steven A., et al. "The extracorporeal life support organization Maastricht treaty for nomenclature in extracorporeal life support. A position paper of the extracorporeal life support organization." American journal of respiratory and critical care medicine 198.4 (2018): 447-451.
John, Kevin J., et al. "Comparison of mixed venous oxygen saturation and premembranous venous oxygen saturation in patients on peripheral venous-arterial extracorporeal membrane oxygenation." ASAIO Journal 70.9 (2024): e130-e132.
Nakano, Kazuya, et al. "Estimation of blood oxygen saturation in the circulation circuit for extracorporeal membrane oxygenation." IEEE Access 7 (2019): 155057-155063.
Mojoli, F., et al. "Temperature monitoring during ECMO: an in vitro study." Intensive Care Medicine Experimental 3.Suppl 1 (2015): A506.
Doo, Irene, et al. "Diagnostic accuracy of infection markers to diagnose infections in neonates and children receiving extracorporeal membrane oxygenation." Frontiers in Pediatrics 9 (2022): 824552.
Walker, Joshua L., et al. "Calculating mixed venous saturation during veno-venous extracorporeal membrane oxygenation." Perfusion 24.5 (2009): 333-339.
Augustin, Simon, et al. "Centrifugal pump inlet pressure site affects measurement." Perfusion 25.5 (2010): 313-320.
Mulholland, J. W., W. Massey, and J. C. Shelton. "Investigation and quantification of the blood trauma caused by the combined dynamic forces experienced during cardiopulmonary bypass." Perfusion 15.6 (2000): 485-494.
Lequier, Laurance, et al. "Extracorporeal membrane oxygenation circuitry." Pediatric Critical Care Medicine 14.5_suppl (2013): S7-S12.
Hermann, Marcus, and Harald Keller. "THE ECMO CIRCUIT AND TROUBLESHOOTING." Extracorporeal Membrane Oxygenation: An Interdisciplinary Problem-Based Learning Approach (2022): 47.
Mulholland, J. W., W. Massey, and J. C. Shelton. "Investigation and quantification of the blood trauma caused by the combined dynamic forces experienced during cardiopulmonary bypass." Perfusion 15.6 (2000): 485-494.
Girod, Grégoire, et al. "Cavitation versus degassing: in vitro study of the microbubble phenomenon observed during echocardiography in patients with mechanical prosthetic cardiac valves." Echocardiography 19.7 (2002): 531-536.
Tanaka, Daizo, et al. "What is the optimal blood pressure on veno-arterial extracorporeal membrane oxygenation? Impact of mean arterial pressure on survival." ASAIO journal 65.4 (2019): 336-341.
Lee, Dong Hun, et al. "Heat loss augmented by extracorporeal circulation is associated with overcooling in cardiac arrest survivors who underwent targeted temperature management." Scientific Reports 12.1 (2022): 6186.
Ortoleva, Jamel P., Abdulaziz Alfadhel, and Adam A. Dalia. "Invasive hemodynamic and physiologic considerations in patients undergoing extracorporeal membrane oxygenation." Journal of Cardiothoracic and Vascular Anesthesia 35.9 (2021): 2549-2551.
Russ, Martin, et al. "Extracorporeal membrane oxygenation blood flow and blood recirculation compromise thermodilution-based measurements of cardiac output." Asaio Journal 68.5 (2022): 721-729.
Rosen, Bradley H. "Circuits, Membranes, and Pumps." Extracorporeal Membrane Oxygenation for Adults. Cham: Springer International Publishing, 2022. 63-79.
Guinot, Pierre-Grégoire, et al. "Passive leg raising can predict fluid responsiveness in patients placed on venovenous extracorporeal membrane oxygenation." Critical Care 15.5 (2011): R216.
Jendoubi, Ali, et al. "Fluid management in adult patients undergoing venoarterial extracorporeal membrane oxygenation: A scoping review." Journal of Critical Care 86 (2025): 155007.