ECMO is half of "Mechanical supports: ECMO/IABP", an L1 topic of Section 2.1.4 in the second edition of the CICM Syllabus for the Second Part Examination. And the ECMO pump seems like a fairly important component to understand. On the other hand, it is also a very basic element, as mostly all it does is spin. How could anybody overcomplicate this subject, even if they wanted?
Well:
Again, as with a lot of this equipment, there did not seem to be a single source of definitive truth which one could recommend to the interested reader. The most valuable insights about the performance characteristics and behaviour of these devices actually came from Purcell & Silvaggio (1997), which is a paper by two mechanical engineers involved in industrial hydraulic projects. Much can be learned from the conversational style and simple clear logic used for their explanations. In contrast, medical literature on this topic trends towards the obscure and weird, where important generalisable points are hidden and scattered across multiple papers, but no single article offers a synthesis of everything in a concise package. Instead of exhorting the reader to absorb all these papers, it is perhaps better to point out the almost total irrelevance of this topic to the CICM exams, and to the understanding of ECMO in general. Many people are satisfied simply knowing that these devices produce flow, and one can operate them without having to grasp the finer points of their design.
The older editions of the ELSO Red Book refer to the centrifugal pump as something towards which there has been "steady shift over the years", perhaps speaking mostly for the roller pump, which is still the dominant cardiopulmonary lifeform worldwide. The peristaltic pump propels blood forward through a length of flexible tubing ("racecourse") which is sequentially compressed along its lengh by rollers on the edge of the wheel. The principle remains essentially unchanged since the days of DeBakey (1934), whose original device is depicted below:

This has several advantages:
However, as one might imagine, the constant application of violence to even the most elastic tubing will eventually cause it to develop stresses and fracture. Kaluza et al (2022), faced with COVID and the possibility of repurposing cardiac theatre equipment, determined that normal PVC tubing tends to disintegrate after only 72 hrs of being gently massaged by the pump rollers. Elders reminiscing on the early phases of ECMO development recall all kinds of nightmares related to having the nurses "walk" the tube each shift, to make sure the tubing is cycled through the pump and no single length ends up being more damaged than the rest. In short, the roller pump is unsuitable for long term use, and this is why ECMO circuits usually feature a centrifugal pump with a magnetically coupled motor.
In function, it resembles the magnetic stirrer one may recall from tedious chemistry practicals, which uses a set of stationary electromagnets with rapidly alternating current to push a ferromagnetic stir bar ("flea", "spinner", etc) in order to mix the contents of the beaker. The pump drive head is suspended magnetically within the chamber and makes no contact with its walls, limiting the sources of red cell destruction to mostly just shear stress. That may sound flimsy, but Schibilsky et al (2020) reported some extremely stiff displacement values - to shift their rotor by 1mm along the Z axis, the investigators recorded needing 57 N of force (much more than is required for cricoid pressure). This is good: there is not much room for the pump to move around; the clearance around the sides of the rotor is only about 2mm on each side, and the total prime volume is usually something like 20-40ml. In other words, the rotor occupies the vast majority of the chamber (the volute, if you want to sound like a pump nerd), and is covered in a fairly thin layer of rapidly moving blood, which is a design compromise between the competing interests of haemodynamic and haemolytic performance. They all have a similar appearance and internal structure. Here is an image to describe the market leaders compared by Han et al (2022):

where a) is the Revolution pump, b) is Rotaflow and c) is CentriMag. And here is an actual photo of a CentriMag device from MacLaren et al, 2012:
.jpg)
The reason they all have a similar design is that all aspects of this device have over years become unsurprisingly overengineered to achieve the highest flow rate with the lowest possible rate of haemolysis. From empirical measurements, we know that the optimal dimensions include an outlet angle of 15–35°, with a blade clearance of around 0.1-0.05mm, a 20° tongue angle and a blade wrap angle of 100–140°with 4-7 blades per pump. All of this, and how we arrived here, are explained in Huang et al (2021), which is also the source of the images below.
.jpg)
The motor behind the pump is a 12-28V DC motor, and it is surprisingly hard to find the technical specifications that describe it. Working backwards from available information, considering it takes the Cardiohelp module around 90 minutes to fully discharge a 10.8 V/6450 mAh (~70 Wh) battery, that makes it a 46.6 Watt motor, approximately equivalent to a small handheld car vacuum or a cordless rotary tool. With that kind of power, it spins a 60-90g pump head at some number of thousands of RPM, where a normal speed would usually be something like 2500-3000 rpm. This is much less than either the aforementioned vacuum or the rotary tool (each could easily go harder than 35,000 rpm), but this is completely reasonable as anything more than 4000 rpm is likely to produce mechanical haemolysis, and the device will not let you crank it harder past 5000 rpm.
The RPM is the variable that is usually in your control. Some pumps may also have various "LPM" modes which allow the user to set a desired flow, which the pump will then try to meet, but the bottom line is that flow is not directly controlled, as it might be in a peristaltic pump. The litres per minute of output is determined by several factors, including the RPMs, but also the "preload" and the "afterload" of the mechanism, i.e. the availability of blood incoming, and the resistance to the outflow.
They are therefore described as "preload-dependent and afterload-sensitive", and this combination of words continues to appear through the ECMO literature, though it is not completely clear what would happen to the meaning of these words if the adjectives were swapped. In any case, this is a genuine feature of the device, and is common to all such centrifugal pumps.
Observe, the data from Clark et al (2011), who tortured a Rotaflow pump with three different afterload settings:

As you can see, per given pump speed RPM setting, the flow was lower with a higher systolic pressure. Why is this so? Moazami et al (2015) is a lucid interpretation of the physics involved, explaining that positive displacement pumps like the peristaltic cardiopulmonary bypass pump or your own heart are completely different. These devices (and organs?) push fixed volumes of blood up the hill, and then leave them there -behind valves or occluded tube lengths, so they cannot roll back. The pump then gets another load to move uphill, and cycles in this fashion ad infinitum. To be sure, eventually the hill gets too steep, and the pump will no longer be able to shift fluid against the counterpressure, but this would be a fair way off. Pasquali (2020) offers a ballpark generalisation that positive displacement pumps drop their flow rate by about 1% for every 100% increase in downstream pressure.
On the other hand, the centrifugal pump is constantly blowing to keep the flow going in the right direction, having no valves to rely on. The movement of blood occurs because the centrifugal pump generates a pressure gradient between the pump end of the circuit and the patient's circulatory system. The result is variable rate of flow, which depends on the resistance in the circuit downstream (which mostly comes from the patient's circulatory system and the cannula diameter). In fact, at low-ish pump speeds, a V-A ECMO patient can overpower the pump with their left ventricle, and generate retrograde flow through their circuit even as the rotor spins against the current. The attentive reader will have noted at this stage that the circuit in the experiments referenced above starts to flow backwards when the systolic pressure is 100 mmHg and the pump speed is below about 1500 RPM.
"Preload dependence" or "preload sensitivity" is even easier to understand, particularly when it is expressed as "nothing left to pump". The most immediate manifestation of this in any circuit is the phenomenon known as "suckdown" in the ICU folk vernacular. One occasionally also sees it referred to as "kicking", "shudder", or "chatter", and probably a dozen other words (chugging? Wooble?) that all basically describe the sudden desire for the ECMO access line to forecefully straighten itself under the effects of atmospheric pressure. It does so when the access pressure becomes extremely negative, and as the result of it the flow across the circuit drops dramatically, which then reduces the negative pressure and allows the line to return to its previous shape. And then this occurs cyclically, usually in association with annoying alarms. Apart from unstable flow, the other disadvantage of this is the cyclical reexposure of blood cells to extremely negative pressure, with all the associated cavitation and outgassing that translate into haemolysis.
Suckdown/chatter/chugging is not the only way preload sensitivity can manifest. The availability of "preload", i.e. sufficient pressure for blood flow to the pump head, is a determinant of how much flow the pump head can generate from a given RPM setting. Working mostly with VADs Salamonsen et al (2011) determined that the sensitivity of these pumps is still less than that of the natural LV, around 0.1 L/min/mmHg at a stable RPM speed (i.e. at their experimental RPM speed the tested pump increased its flow rate by 100ml/min for every 1mmHg increase in preload pressure). In contrast, devices that operate by volume displacement (eg. peristaltic pumps and various syringe or piston-based devices) tend to be indifferent to preload, and will continue to produce the same volume per minute until the reservoir supplying them has literally run dry.
Schibilsky, David, et al. "Hemocompatibility of new magnetically-levitated centrifugal pump technology compared to the CentriMag adult pump." Scientific Reports 10.1 (2020): 22055.
Hijikata, Wataru, et al. "A magnetically levitated centrifugal blood pump with a simple‐structured disposable pump head." Artificial organs 32.7 (2008): 531-540.
Huang, Bo, et al. "Geometric optimization of an extracorporeal centrifugal blood pump with an unshrouded impeller concerning both hydraulic performance and shear stress." Processes 9.7 (2021): 1211.
MacLaren, Graeme, Alain Combes, and Robert H. Bartlett. "Contemporary extracorporeal membrane oxygenation for adult respiratory failure: life support in the new era." Intensive care medicine 38.2 (2012): 210-220.
DeBakey, Micheal E. "A simple continuous flow blood infusion instrument." New Orleans Med Surg J 87 (1934): 386.
Guihaire, Julien, et al. "Physiology of the assisted circulation in cardiogenic shock: a state-of-the-art perspective." Canadian Journal of Cardiology 36.2 (2020): 170-183.
Moazami, Nader, et al. "Does pulsatility matter in the era of continuous-flow blood pumps?." The Journal of heart and lung transplantation 34.8 (2015): 999-1004.
Han, Dong, et al. "Computational fluid dynamics analysis and experimental hemolytic performance of three clinical centrifugal blood pumps: Revolution, Rotaflow and CentriMag." Medicine in novel technology and devices 15 (2022): 100153.
Clark, Joseph B., et al. "An investigational study of minimum rotational pump speed to avoid retrograde flow in three centrifugal blood pumps in a pediatric extracorporeal life support model." Perfusion 26.3 (2011): 185-190.
Salamonsen, Robert F., David G. Mason, and Peter J. Ayre. "Response of rotary blood pumps to changes in preload and afterload at a fixed speed setting are unphysiological when compared with the natural heart." Artificial organs 35.3 (2011): E47-E53.
Purcell, John E., and Joseph A. Silvaggio. "A comparison of positive displacement and centrifugal pump applications." (1997).