"Sweep" is the colloquial term usually used to refer to the gas that blows through the hollow fibres of the ECMO oxygenator, supplied by the air/oxygen gas blender. ECMO is half of "Mechanical supports: ECMO/IABP", an L1 topic of Section 2.1.4 in the second edition of the CICM Syllabus for the Second Part Examination, and the gas blender is the least technologically advanced component of the system. It is a crude mechanical device composed of machined steel plates and bolted together in a manner which resembles 1970s Soviet cars, a totally analogue steampunk anachronism among the high-tech equipment of the ICU.
In short:
The local model is supplied by Sechrist, who also manufacture hyperbaric oxygen chambers, and their site has informative user manuals, but that's the extent of the literature about these things.
Those who enjoy assembling maddeningly complex Ikea furniture will also be able to appreciate the beauty of the Sechrist service manual for the 3500 module, which contains such excellent diagrams as this one:

This diagram is reproduced here with the caveat that it is purely decorative. If the process of troubleshooting the ECMO circuit has lead you to a point where you are convinced that you need to to disassemble this device, then the problem is almost certainly you, and not the device. Put the screwdriver away. For basic operation it is probably only necessary to know where the gas goes in and comes out. The blender functions as a proportioning valve, where a full turn in one direction changes the proportion of the gas flow to 100% of that input. The valve dial simply allows you to adjust the air/O2 proportions with percentage values. The flow regulator is then used to adjust the flow, which is the actual proportion of the supplied gas.

The tubing attaching the gas blender to the oxygenator is a familiar green colour to identify it, and it happens to be 1/4 inch tubing (rather than the usual 5mm 3/16th tubing) for reasons which are obscure and probably largely historical. Certainly nothing about the gas flowing in the extra-thick ECMO gas tube is different from the gases that flow through the regular crush-resistant oxygen tube, so it is not because of some physics-related reason. The only possible answer that presents itself is the need to clearly differentiate the ECMO gas supply from the regular oxygen supply so that they cannot be accidentally exchanged (and the ECMO circuit is blasted with 15L/min while the norebreather mask remains inexplicably flat). In other words, it may be a safety feature, but this is the author's personal speculation, and there is certainly nothing in any of the literature to explain this specific design choice.
An additional safety feature is the water and particle trap at the back of the device, which keeps terrible things from blowing randomly out of the 1950s-era air pipes of your hospital, and into the oxygenator.

The mixed gas passes through one or two flow regulators; a model often used allows both fine and coarse control of the flow, where one regular adjusts in increments measured L/min and the other 100ml/min. The flow regulators are in parallel rather than in series, which means they are additive - total flow is combined at the end into a single gas stream, which means if you dial 500ml/min on one regulator and 2L/min on the other one, the total flow is 2500 ml/min.

So, why do you even need a separate device for this? The answer is "you don't". De Roux et al (2020) established that in a resource crisis you can simply do the maths yourself, and hook the oxygenator directly up the wall outlets via a standard Y-connector. Moreover, a common practice is to just crank the oxygen up to 100%, i.e. no "blending" of any sort is happening anyway. Whether it is better to keep it that way, or whether we should protect patients from hyperoxia, is the subject of the BLENDER trial.
"Sweep" has become the term for the bulk flow of gas in the oxygenator, borrowed from the pages of chemistry textbooks, where "sweep gas" described a carrier gas or vapour used to test permeability of membranes or to separate gas mixtures. The first reference to this use comes from Maier (1940), who further attributes the original invention to a 1923 paper by Schrader, Jaeger and Fischer (the famous Franz Fischer of the Fischer-Tropsche process we all know and love). It is unfortunately lost in a distant paper copy of Brennstoff-Chemie, and so we do not know what the original-language wording was (modern German literature has come to use Spülgas, which translates better as "purge gas").
In essence, where a membrane is being tested for permeability (or being used to selectively remove a gas from a mixture), where the test gas mixture is on one side of the membrane, the other side is exposed to a constant flow of the sweep gas, so that a constant concentration gradient for diffusion is sustained. The sweep gas is then carried to a detector (to assess the amount of test gas that has made its way through the membrane, as an evaluation of its permeability) or just purged if all you are interested is separating the gas mixture. One can immediately see that this is precisely what happens in the ECMO oxygenator: a membrane separates two gas mixtures. All the early developers of oxygenators (eg. Clowes Jr et al , 1956, or Gibbon Jr in 1967) were not chemists and did not have this language, but when engineers became involved in the design of the such membranes (eg., Mathewson & Ryan, 1971), they tested their membranes for permeability using O2 as the sweep gas, and referred to it as such in their reports. Ergo, by the mid-1970s, the use of "sweep" to describe oxygenator gas had crept into the lexicon of perfusionists and cardiac anaesthetists, such as Nash et al (1977) and Mandl (1977), along with the novel membranes.
But that's irrelevant, one might correctly complain, having started reading a section subtitled "Sweep settings for ECMO" and having thus far encountered zero discussion of the sweep settings. Yes, let's refocus on the pragmatic needs of setting this device. In short, the amount of gas flow required by the patient is going to be fairly individual, but a good ballpark estimate is to match the flow of blood, 1:1. In other words, if the ECMO flow rate is 5L/min, so should the gas flow rate be 5L/min. The precise adjustments can then be made on the basis of serial gases. If all other factors remain equal, the changes in sweep gas flow will function similarly to changing the minute volume on the ventilator, i.e. there will be a fall in PaCO2 with higher sweep; but all things are never in equilibrium and the flow may need to be adjusted up and down as variables are altered (eg. the performance of the patient's own circulation and pulmonary gas exchange).
It is hard to come up with something the blender might do to bedevil your night shift, as it is usually a very reliable member of the extracorporeal life support team. Possible blender misbehavioiurs include:
De Roux, Quentin, et al. "A simple approach for gas blender on extracorporeal membrane oxygenation in resource shortage context." ASAIO Journal 66.10 (2020): 1076-1078.
Burrell, Aidan, et al. "Blend to limit OxygEN in ECMO: a RanDomised ControllEd Registry (BLENDER) trial: study protocol and statistical analysis plan." Critical Care and Resuscitation 25.3 (2023): 118-125.
FISCHER, F., SCHRADER, H., and JAEGER, A. McGraw Hill Co., New York, Über die Trennung von Gasgemischen durch Diffusion bei Wasserdampfgegendruck. Brennstoff- Chemie, vol. 4, 1923, p. 289
Maier, Charles George. Mechanical Concentration of Gases. Vol. 431. US Government Printing Office, 1940.
Mathewson, Wilfred F., and David M. Ryon. "Membrane Requirements in the Pursuit of Blood Oxygenator Optimization." Membrane Processes in Industry and Biomedicine: Proceedings of a Symposium held at the 160th National Meeting of the American Chemical Society, under the sponsorship of the Division of Industrial and Engineering Chemistry, Chicago, Illinois, September 16 and 17, 1970. Boston, MA: Springer US, 1971.
Clowes Jr, George HA, Amos L. Hopkins, and William E. Neville. "An artificial lung dependent upon diffusion of oxygen and carbon dioxide through plastic membranes." Journal of Thoracic Surgery 32.5 (1956): 630-637.