Transcutaneous pacing is a mechanism to produce chest wall twitching and small rectangular burns on the surface of your patient, associated with the occasional risk of clinically useful cardiac depolarisation. Though cardiac pacing is a Level 1 topic from Section 2.1.19 in the first edition of the CICM Syllabus for the Second Part Examination, on a literal reading of the syllabus the transcutaneous option could legitimately be dismissed by the reader as "not an ICU procedure". And indeed it is reasonable to view it as such, considering that the majority of transcutaneous pacing in the world is done pre-hospital, and the intensivist is usually involved at the point where the time has come to convert to transvenous. Still, it is very worth while knowing all about this technique, as such patients would inevitably be referred to ICU wherever they appear. Past paper questions on this topic are few: only Question 1 from the first paper of 2024 asked about it in any great detail. In summary:
Of the learning objectives listed for cardiac pacing in general ("for each of the above expected knowledge will include"), the indications, relevant anatomy, insertion procedure for transvenous wires, advantages and disadvantages (including different modes) and all the contraindications or complications are listed elsewhere, and this section mainly covers the expected knowledge for transcutaneous pacing specifically. In terms of peer-reviewed reading, Doukky et al (2019) is probably the single most useful reference for the sorts of things that one might find in CICM exams.
If your argument against chronotropes was that "this is not a sustainable solution", then surely the same argument is even more true for transcutaneous pacing. It is hard to warm to this therapy, as it has all the rudeness of painful electrical shocks without any of the charm of effective cardiac output enhancement. Since it was first described by Paul Zoll in 1952, the technique has only really changed by incorporating better electrode technology. The authors used subcutaneous needles to pass a current across the ventricles of two patients in "ventricular standstill"; both would have survived if one had not developed cardiac tamponade from "34 intracardiac injections of epinephrine". A revision by Falk et al (1983), who was experimenting on his faculty colleagues and medical students, had used "large electrodes moistened with tap water". Determined to make the best of it, Doukky et al (2019) recommend the following modern technique:

"Capture", in the context of the external pacing scenario, does not mean "stare at the monitor to observe for characteristic broad complexes", as the complexes are already likely to be idioventricular and abnormal, and moreover there will usually be abundant musculoskeletal artifact. This pacing artifact may mimic a QRS complex, and will confuse the unready. One may be able to distinguish them: the pacing artifact is usually narrower and solitary, whereas a "proper" captured QRS is wide, resembling left bundle branch block morphology, and followed by an angry-looking T wave. Additionally, the device tries to fool you by detecting its own residual current flow and reporting it as a "false capture" artifact that can appear in the right position to look like a peaky T-wave. Here, a couple of examples shamelessly stolen from the excellent blog entry at aclsmedicaltraining.com demonstrate the key features: 
And even when the morphology of the complexes looks correct, technically speaking there is no reason for this not to be PEA, insofar as a myocardium will still confidently produce these complexes even as it stands completely empty (or full of clot). Therefore, logically, it would make sense to distrust the electrical information, and one should either feel the pulse or observe the arterial waveform to confirm that each shock is indeed followed by a myocardial response.
How much current? User experience may vary. Remembering that Ohm's law describes the relationship V = IR, and that the usual resistance across a normally sized dry adult chest varies from 12 to 200 Ω (Heyer et al, 2022), one should be surprised to have one current fit all occasions, and it would be reasonable to expect it to vary by more than a factor of ten from person to person. The usual range of values given by most papers is 20-140 mA, and all of them for some reason quote the original paper by Falk and the Zolls from 1983.
Is that a lot? The medical professional rarely encounters electricity measured in this way. A comparison with external defibrillation is somewhat confused by the fact that one typically defibrillates with Joules of energy, but one paces with Amperes of current. The main reason for this is that, though in both defibrillation and pacing the delivered current is the key, for defibrillation to be effective one needs to deliver an amount of current necessary to depolarise a critical mass of myocardium (perhaps 50% of the mass), for which an adult 400g heart requires a dose of around 1 joule of energy (Geddes et al, 1974). To achieve this kind of energy transfer, one needs to apply a considerably larger current externally, expecting a large loss of energy into surrounding structures (particularly the poorly conductive skin). Empirical measurements have determined that this is about 200J for a biphasic shock and about 300J for a monophasic one. The defibrillator is therefore set to deliver this known amount of energy into an object of unknown properties. The voltage being a fixed operating characteristic of the capacitor inside the machine (2,700 V for the Zoll M series), the remaining variables are the current and the impedance. In order to achieve the desired energy transfer over the duration of the standardised pulse (6-12 milliseconds), the defibrillator tests the impedance between the skin electrodes to determine the necessary current, which ((I = V/R) would be for 27A for a 100 Ω human torso.
In contrast, to pace the myocardium one merely needs to depolarise a small amount of tissue, from where the action potentials will then spread normally. The current required is therefore much smaller. The impedance being stable between beats and pulse duration being fixed (100Ω and 20-40 ms for pacing), one sets the current, and the remaining property is the voltage, which is then manipulated to adjust Ohmian V = IR relationship. For a 100mA current through 100Ω that would be about 10V, which (delivered over 40 milliseconds) gives an energy of 0.04 J per pulse.
So, 0.04 J per pulse is 3.6 J for a full minute of pacing at a rate of 90, or 216 J for an hour. A person who has only ever experienced zero Joules of electricity may have trouble contextualising this information. How much does 3.6 Joules hurt? Is sedation mandatory? One of Zoll's patients from 1952 remained conversant throughout the period of being paced through needles, though in all fairness he was probably too busy experiencing "incapacitating angina pectoris" to notice. Of Falk's cohort, two wusses "requested discontinuation of stimulation at 60 mA because of discomfort". Klein et al (1988) attempted a courageous exploration of pain thresholds in consenting volunteers and determined that fully half of them could not tolerate even to have the threshold determined, let alone a sustained period of effective capture. Bektas & Soyuncu (2016) reported that it "is tolerated by conscious patients despite the painful intervention" but then confessed that some of their 91-patient cohort had required enough sedation to affect their vital signs. That analgesia is necessary for humanitarian reasons is also relatable to the authors's personal experiences arising from uneducated guesses about the misbehaviour of his domestic wiring. So, in summary, this technique requires sedation and analgesia in a patient who - if they absolutely need transcutaneous pacing and can't get a wire - is by definition unstable.
Other disadvantages, as if we need to list any more, include:
Why, then, would you do this to anyone? Well:
How temporary is it? Or, rather: how long can one persevere with transcutaneous pacing? Sources that claim "transcutaneous pacing has been used continuously for as long as 4-5 days" typically quote ancient manuscripts by Falk et al (1983) and Zoll et al (1985), but even these pioneers do not actually report such data: Falk et al reported only the safety of 15-30 minute periods of pacing, and the Zolls mentioned only "prolonged intervals up to 25 hrs". Reported durations of longer pacing sessions actually take the form of cautionary tales, eg. where Payne et al (2018) and Carrizales-Sepúlveda et al (2018) reported burns following 12-36 hours of pacing. The authors attributed these effects to the dessication of the pads, resulting in uneven current distribution, and recommended that the pads be changed on a regular basis (4-6 hourly). The reader should leave with the distinct impression that to pace somebody transcutaneously for longer than a single digit number of hours would take them deep into case report territory from the 1980s.
Bektas, Firat, and Secgin Soyuncu. "The efficacy of transcutaneous cardiac pacing in ED." The American journal of emergency medicine 34.11 (2016): 2090-2093.
Doukky, Rami, et al. "Using transcutaneous cardiac pacing to best advantage: How to ensure successful capture and avoid complications." The Journal of critical illness 18.5 (2003): 219.
Le, Khoi, and Nora Goldschlager. "Temporary cardiac pacing in the intensive care unit." Journal of Intensive Care Medicine 11.2 (1996): 57-78.
Verbeet, Thierry, José Castro, and Pierre Decoodt. "Transesophageal pacing: a versatile diagnostic and therapeutic tool." Indian Pacing and Electrophysiology Journal 3.4 (2003): 202.
Falk, Rodney H., Paul M. Zoll, and Ross H. Zoll. "Safety and efficacy of noninvasive cardiac pacing. A preliminary report." The New England journal of medicine 309.19 (1983): 1166-1168.
Zoll, Paul M. "Resuscitation of the heart in ventricular standstill by external electric stimulation." New England Journal of Medicine 247.20 (1952): 768-771.
Zoll, Paul M., et al. "External noninvasive temporary cardiac pacing: clinical trials." Circulation 71.5 (1985): 937-944.
Moayedi, Siamak, et al. "Anteroposterior pacer pad position is more likely to capture than anterolateral for transcutaneous cardiac pacing." Circulation 146.14 (2022): 1103-1104.
Heyer, Yasmine, Daniela Baumgartner, and Christian Baumgartner. "A systematic review of the transthoracic impedance during cardiac defibrillation." Sensors 22.7 (2022): 2808.
Syverud, Scott A., et al. "Hemodynamics of transcutaneous cardiac pacing." The American Journal of Emergency Medicine 4.1 (1986): 17-20.
Dillon, Stephen M., and Kevin F. Kwaku. "Progressive depolarization: a unified hypothesis for defibrillation and fibrillation induction by shocks." Journal of cardiovascular electrophysiology 9.5 (1998): 529-552.
Geddes, L. A., et al. "The electrical dose for ventricular defibrillation with electrodes applied directly to the heart." The Journal of Thoracic and Cardiovascular Surgery 68.4 (1974): 593-602.
Tacker Jr, W. A., et al. "Energy dosage for human trans-chest electrical ventricular defibrillation." New England Journal of Medicine 290.4 (1974): 214-215.
Klein, Lawrence S., et al. "Transcutaneous pacing: patient tolerance, strength-interval relations and feasibility for programmed electrical stimulation." The American Journal of Cardiology 62.16 (1988): 1126-1129.
Smida, Tanner, et al. "Prehospital transcutaneous cardiac pacing in the United States: treatment epidemiology, predictors of treatment failure, and associated outcomes." Prehospital Emergency Care (2024): 1-8.
Kimbrell, Joshua, et al. "False electrical capture in prehospital transcutaneous pacing by paramedics: a case series." Prehospital Emergency Care (2024): 1-9.
Morrison, Laurie J., et al. "A randomized controlled feasibility trial comparing safety and effectiveness of prehospital pacing versus conventional treatment:‘PrePACE’." Resuscitation 76.3 (2008): 341-349.
Mishra, Roly, Parna Thakkar, and Hemant Mehta. "Transcutaneous pacing: a life saviour." International Journal of Research in Medical Sciences 8.6 (2020): 2332.
Carrizales-Sepúlveda, Edgar Francisco, et al. "Thermal burn resulting from prolonged transcutaneous pacing in a patient with complete heart block." The American Journal of Emergency Medicine 36.8 (2018): 1523-e5.
Payne, Joshua E., et al. "Third-degree burns associated with transcutaneous pacing." Case Reports 2018 (2018): bcr-2018.
Altamura, Giuliano, et al. "Transcutaneous cardiac pacing for termination of tachyarrhythmias." Pacing and Clinical Electrophysiology 13.12 (1990): 2026-2030.